ZM4 Instruction Manual 9R87-B June 2012
26
Volts/mA Auto Ramp Setup
Figure 21: Volts/mA Auto Ramp Setup Display
Reference Function
Start Select to start the Auto Ramp simulation.
Mode: Volts/mA mode for the simulation. For
reference only, cannot be modified here.
Begin Value: Starting simulation value. Brings up edit
display to change the value.
End Value: Ending simulation value. Brings up edit
display to change the value.
Ramp Time: Time to ramp between the Begin and End
values. Brings up edit display to adjust the
value between 1 and 300 seconds.
Startup Delay:
Delay after pressing Start key before
simulation begins. Brings up edit menu to
select the value between None, 30 or 60
seconds.